Principle of the Test
All laboratory assays require internal quality control (QC) as a component of the laboratory’s Quality Assurance program. Control ranges are determined with each lot change of reagent, with each lot change of control, and with major service of the instrument. Thereafter controls are run according to the requirements of the Clinical Laboratory Improvement Amendments. Currently the minimum requirement is two levels of control run every 24 hours; laboratories should consult the current CLIA or ISO regulations and the reagent manufacturers’ recommendations when they design their internal QC programs.
PlasmaCon LA control plasma is manufactured from a blend of normal and LA positive plasmas to which buffers and stabilizers have been added, and is used as an LA positive control for lupus anticoagulant tests such as LupoTek KCT, LupoTek Detectin VL, and LupoTek Correctin VL.